General Properties
•
Techniques
•
Delivery Set
•
Specifications
Using up-to-date and high-quality computer technologies our
company produced ECG system
Poly-Spectrum-12
which combines the unique diagnostic potentialities and
surprisingly low price.
Poly-Spectrum-12
General Properties
·
High quality recording from 1 to 12 ECG leads
·
ECG recording in Frank orthogonal leads system
·
ECG measurement and interpretation
·
Printing ECG on plain paper by jet or laser printer
·
A book in size ECG unit
·
USB interface with the computer
·
Optional software modules:
o
Poly-Spectrum-Rhythm
for heart rate variability analysis
o
Poly-Spectrum-QT
for Q-T interval dispersion analysis
o
Poly-Spectrum-VLP
for ventricular late potentials registration and analysis
o
Poly-Spectrum-Ergo
for exercise testing
·
Built-in phono- and sphygmochannels (The unit can be
supplemented with the phonocardiogram and sphygmogram
registration module Poly-Spectrum-PhS)
Poly-Spectrum-12
Techniques
|
|
Patient B.
12-channel ECG on a computer screen. A user can change
leads system (Frank, Nehb), number of displayed leads,
paper speed, sensitivity and set of filters. |
|
|
|
|
|
Patient B.
ECG measurement and interpretation. Selected by a user
representative or averaged QRS-complex is displayed in
the window. Markers are installed in the QRS-complex
reference points. At the bottom of the window the area
with automatically generated interpretation is located. |
|
|
|
|
|
Patient B.
Clicking on any lead, you can zoom this QRS-complex for
more details. You can move a marker in the window, if
you think the program located it incorrectly. |
|
|
|
|
|
Patient B.
The table of all measurements. |
|
|
|
|
|
|
Samples of checkup reports, automatically generated by
Poly-Spectrum-12 software.
|
|
|
|
|
|
Poly-Spectrum-12 Delivery Set
·
ECG electronic unit with a stand
·
Patient cable
·
Reusable limb clamp ECG electrode – 4 pcs.
·
Reusable prеcardial suction chest ECG electrode – 6 pcs.
·
Electrode gel
·
CD with software
·
User and technical manuals
·
Transportation bag |
|
Poly-Spectrum-12
Specifications
|
Number of ECG channels |
12 |
|
ECG leads |
I, II, III, aVR, aVL, aVF, V1…V6; X, Y, Z (by
Frank) |
|
Number of breath channels |
1 |
|
A/D converter |
24 bit |
|
Sampling rate |
1000 Hz |
|
Frequency response |
0 - 250 Hz |
|
Low pass filter |
35 Hz (myogram), 75 Hz |
|
High pass filter |
0.05 Hz (3.2 s) |
|
AC filter |
50 (60) Hz |
|
Drift filter |
Yes |
|
Sensitivity |
2.5, 5, 10, 20, 40 mm/mV |
|
Paper speed |
5, 10, 25, 50, 75, 100, 200 mm/s |
|
Common-mode rejection ratio, not less than |
100 dB |
|
Input range |
1µV-330 mV |
|
Input impedance, not less than |
50 MΩ |
|
Input noise level, not more than |
5 µV |
|
Defibrillation protection |
Yes |
|
Control of electrode setting |
Yes |
|
Interface |
USB |
|
Supply Voltage: |
|
ECG unit |
5 V DC |
|
Desktop PC-based system |
220/230 V AC (50 Hz)/110 V AC (60 Hz) |
|
Notebook PC-based system |
220/230 V AC (50 Hz)/110 V AC (60 Hz) / Int.
battery |
|
ECG unit power consumption, not more than |
1.2 VA |
|
Dimensions: |
|
ECG unit |
150x200x60 mm |
|
Delivery set |
450x350x160 mm |
|
Weight: |
|
ECG unit (with cable and electrodes) |
1.2 kg |
|
Delivery set (without computer and printer), not
more than |
10 kg |
|
Safety |
BF type |
|
|